ECRHS Nebuliser Calibration Phase (1)

Conditions:

  • Static driving pressure 180 kPa +/- 5%.
  • Activation time 1 second.
  • See also: Graph of Results

Nebuliser

Aerosol output (mg)

101

6.5

102

6.8

103

6.7

104

6.5

105

6.7

106

7.4

107

7.3

108

7.4

109

6.5

110

7.4

111

7.4

112

7.2

113

7.4

114

5.3

115

7

116

7.1

117

7.2

118

7

119

6.4

120

7.2

121

7.2

122

7.3

123

4.4

124

7.1

125

7.2

126

7

127

6.8

128

7.6

129

5.8

130

7.2

131

4

132

7.8

133

7

134

4.6

135

6.7

136

8

137

7.8

138

7.3

139

7.1

140

7.2

141

7.6

142

7.9

143

7.6

144

7.6

145

7.2

146

7.3

147

5.8

148

7.3

149

7.2

150

7.3

151

7.4

152

6.7

153

7.2

154

7.2

155

7.4

156

7.1

157

7.8

158

7.4

159

7.9

160

7.5

161

8.3

162

7.5

 

 

Note: nebulisers marked in yellow

 

have been excluded due to low output.

 