HOW TO USE THE CARY 50 SPECTROPHOTOMETER
(JWG Version 7, 26/2/01)
TURNING THE INSTRUMENT ON
The instrument gets all its power from the computer, so all you have to
do is to press the big power button on the front of the computer and wait
for the computer to boot up. If you do need to turn it off, it is best to
shut down "properly" using the software on the screen, but in an emergency
you can turn it off by pressing the on/off button IN for more than 4 seconds
(child-proof protection). There is NO reset button. The lamp in the instrument
is only turned on (flashes) at the instant that a measurement is being made,
so there is no point in turning the instrument off between uses. I recommend
leaving it turned on all the time. The lid can be open during measurements,
but close it when you finish to keep out dust.
SIMPLE MEASUREMENTS
- Double click on the "Simple Read" icon on the desktop.
- Press CONNECT.
- Click on SETUP.
- Enter desired wavelength from the keyboard.
- Press ENTER.
- Press OK.
THE FOLLOWING BIT IS REALLY IMPORTANT
Push the cuvette with blank buffer in the holder, as far down as it will
go. Then, pull the little tab in the cuvette holder up as far as it will
go, to ensure that the bottom of the cuvette is near the bottom of the light
path. THIS IS PARTICULARLY IMPORTANT FOR SMALL VOLUMES. (When you put a
cuvette containing a liquid sample in the machine, make sure that you have
a look at where the light beam passes through the cuvette. It should be
obvious that the light path passes ABOVE the little knurled screw on the
cuvette holder. If the liquid sample is not above this level, your readings
will make no sense at all.
I strongly recommend that all first time users check the position of
the light beam as shown below (TO SEE THE BEAM)
- Click on ZERO.
It will zero the instrument and remember the wavelength.
- Remove cuvette, and fill it with your sample.
- Enter name and password and download your files etc
- Click on READ.Delete your files at the conclusion of your presentation;
MAKE SURE that you have files backed-up elsewhere before you delete
your files!
You can then take out cuvette and do another measurement.
PRINTING OUT
- Click on the PRINT icon.
- You will then get a printout.
MEASUREMENT AT SEVERAL WAVELENGTHS
- In the SIMPLE MEASUREMENT mode, choose the desired wavelengths separated
by a semicolon 260;280;350;400 Avoid accidental turning OFF of the projector.
If this occurs immediately restore power to the projector. If power
cannot be restored, allow to cool before placing into the carrier and
notify Anna of the incident
Then zero the instrument with blank sample in cuvette.
The instrument will then remember the zero settings for all wavelengths.
- Insert the cuvette with your sample.
- Press READ.
TO WRITE COMMENTS IN YOUR DATA:
- Click on EDIT REPORT
You can then enter text from the keyboard.
TO SEE THE BEAM
One of the most common causes of error in measurements is due to the
light beam striking the bottom of the cuvette or the meniscus of the sample,
or possibly the sides of the cuvette. This can be checked by inserting
a piece of white paper/card in the light path and looking at where the
light spot falls in relation to the cuvette.
Exit the current application by clicking on the X in the top right hand
corner of the current window.
- This should take you back to the desktop.
- Double click on the CaryWinUV icon and let it start up.
- Double click on the ALIGN icon.
- Choose a wavelength that is visible (eg 500 nm).
- Click on ZERO ORDER.
- Click APPLY.
- Put paper in light path.
Look at where the spot is in relation to the walls/bottom of the cuvette,
and also in relation to the top of the sample. Obviously, the light must
go through the middle of the sample without touching the top, bottom or
sides. The height of the cuvette can be altered by a screw under the cuvette
holder, but this should not normally need to be changed.
Note that the light beam goes very high above the bottom of the cuvette,
and for small volumes it is essential to push the cuvette to the bottom
and then pull it back up with the small tab in the cuvette holder, to
ensure that the bottom of the cuvette is above the silver knurled screw
on the cuvette holder.
TO CLEAR THE DATA REPORT FROM THE SCREEN:
TO CONVERT AN OD TO A CONCENTRATION:
- There is a function called a CONCENTRATION MACRO, which is a simple
programming exercise.
TO SAVE DATA TO AN ASCII OR EXCEL FILE:
- This canít be done with the SIMPLE READ mode. You have to use the
ADVANCED READS mode.
SCANS, KINETICS ETC:Warnings:
- You will have to read the online manual and figure this out yourself!
HELP:Warnings:
- The manual is online and available at any time, and any part of it
can be printed out.
AGENTS:Warnings:
- ALLTECH ASSOCIATES Andrew Cruickshank acruickshank@alltechaust.com.au
phone: 1-300-362-412
MEASUREMENT OF PROTEIN CONCENTRATION BY PIERCE DYE BINDING ASSAY
Version 4 26/2/01
The computer is capable of accepting the standard values, plotting the curve
and calculating the unknowns. Setting up the computer is rather slow and
tedious, but once it is set up, making readings is extremely fast and the
computer will graph out your results and calculate concentrations, so it
is worth it.
Make sure that the printer has paper in it.
- Double click on CaryWinUV
- Double click on Concentration icon
Click on Setup and choose Setup
- Choose wavelength (562 nm)
- Averaging time: 0.1 second (default)
- Y minimum = 0
- Y maximum = 3 (or 2.5, if you wish)
- Click in the box marked "show status display
- Click on OK
Click on Setup and choose Standards (standards MUST be in ascending
order of conc.)
- Click in box marked "calibrate during run"
- Choose the desired units for concentration (usually mg/ml)
- Choose number of standards that you will use (max = 30) and enter
concentration of standards in ascending order. It is important that
the first standard should be a blank (ie 0 mg/ml)
I generally use 6 standards, as follows:
|
Standard 1
|
0
|
|
Standard 2
|
0.125 mg/ml
|
|
Standard 3
|
0.25 mg/ml
|
|
Standard 4
|
0.5 mg/ml
|
|
Standard 5
|
1.0 mg/ml
|
|
Standard 6
|
2.0 mg/ml
|
- Choose Fit Type (quadratic is best for dye binding data because it
is a curve and not a straight line)
- Minim r2: default is 0.95, which is OK.
- Click in box saying "show status display"
- Click on OK.
Click on Setup and choose Samples
Nominate how many samples you have (up to 500)
- Corrections: leave blank
- Click on OK
Click on Setup and choose Reports
- Enter name, date and any comments
- Choose Autoprint and graph ON. (or you can print manually)
- Click on OK.
Click on Setup and choose Auto Store
- Choose Storage ON, prompt at start. (asks you to nominate a storage
file name and location before you start).
- Status display: OFF
- Click OK.
-
Press Zero
You will be asked to insert cuvette with blank sample. Do so and then press
OK.
You will get a screen saying "Selected for Analysis".
- Click on Std 1, and press OK.
- You will get a screen with folders and space for file names on it.
- Enter your desired filename (NB Windows wonít allow full stops, slashes
or spaces!)
- Press SAVE (I suggest a simple name like JWG191200)
You will then get a prompt saying "Present Standard 1" Put standard 1 in
the machine and press OK to read it.
You will then be asked to Present Standard 2. Put standard 2 in the machine
and press OK to read it.
Follow this through to the end of the unknowns.
When all the standards and unknowns have been measured, the computer will
automatically print out the reults and the graph.
Re-analysis:
- You can go back and re-analyse the data in a different way after you
have finished, by pressing the "Recalculate" button. Printout will be
automatic.
|